Prior to performing the cell measurements you should perform all three of the calibration measurements. Please do not proceed unless you have completed the three calibration protocols. Completion of the calibrations will ensure that you understand the measurement process and that you can take the cell measurements under the same conditions. For the sake of consistency and reproducibility, we are requiring all teams to use E. coli K-12 DH5-alpha. If you do not have access to this strain, you can request streaks of the transformed devices from another team near you, and this can count as a collaboration as long as it is appropriately documented on both teams' wikis. If you are absolutely unable to obtain the DH5-alpha strain, you may still participate in the InterLab study by contacting the Measurement Committee (measurement at igem dot org) to discuss your situation.
For all of these cell measurements, you must use the same plates and volumes that you used in your calibration protocol. You must also use the same settings (e.g., filters or excitation and emission wavelengths) that you used in your calibration measurements. If you do not use the same plates, volumes, and settings, the measurements will not be valid.
- E. coli DH5 alpha
- LB Broth+chloramphenicol
- chloramphenicol
- Negative control
- Positive control
- Test Device 1
- Test Device 2
- Test Device 3
- Test Device 4
- Test Device 5
- Test Device 6
baseline absorbance of culture (day 2) measurements of culture (day 2)
0 hr absorbance timepoint measurements of plate 1
0 hr fluorescence timepoint measurements of plate 1
6 hr absorbance timepoint measurements of plate 1
6 hr fluorescence timepoint measurements of plate 1
6 hr absorbance timepoint measurements of plate 2
6 hr fluorescence timepoint measurements of plate 2
- Transform
Negative control
DNA intoE. coli
DH5 alpha
and plate transformants on LB Broth+chloramphenicol. Repeat for the remaining transformant DNA:Positive control
,Test Device 1
,Test Device 2
,Test Device 3
,Test Device 4
,Test Device 5
, andTest Device 6
. - Provision 16 x culture tubes to contain
culture (day 1)
- Inoculate
E. coli
DH5 alpha+Negative control transformant
into 5.0 milliliter of LB Broth+chloramphenicol in culture (day 1) and grow for 16.0 hour at 37.0 degree Celsius and 220.0 rpm. Repeat this procedure for the other inocula:E. coli
DH5 alpha+Positive control transformant
,E. coli
DH5 alpha+Test Device 1 transformant
,E. coli
DH5 alpha+Test Device 2 transformant
,E. coli
DH5 alpha+Test Device 3 transformant
,E. coli
DH5 alpha+Test Device 4 transformant
,E. coli
DH5 alpha+Test Device 5 transformant
, andE. coli
DH5 alpha+Test Device 6 transformant
. Inoculate 2 replicates for each transformant, for a total of 16 cultures. - Provision 16 x culture tubes to contain
culture (day 2)
- Dilute each of 16
culture (day 1)
samples with LB Broth+chloramphenicol into the culture tube at a 1:10 ratio and final volume of 5.0 milliliter. Maintain at 4.0 degree Celsius while performing dilutions. - Measure baseline absorbance of culture (day 2) of
culture (day 2)
at 600.0 nanometer. - Provision 16 x 50 ml conical tubes to contain
back-diluted culture
The conical tube should be opaque, amber-colored, or covered with foil. - Back-dilute each of 16
culture (day 2)
samples to a target OD of 0.02 using LB Broth+chloramphenicol as diluent to a final volume of 12.0 milliliter. Maintain at 4.0 degree Celsius while performing dilutions.
- Provision 16 x 1.5 mL microfuge tubes to contain
cultures (0 hr timepoint)
- Provision a 96 well microplate to contain
plate 1
- Hold
cultures (0 hr timepoint)
at 4.0 degree Celsius. This will prevent cell growth while transferring samples. - Hold
plate 1
at 4.0 degree Celsius. - Transfer 1.0 milliliter of each of 16
back-diluted culture
samples to 1.5 mL microfuge tube containers to contain a total of 16cultures (0 hr timepoint)
samples. Maintain at 4.0 degree Celsius during transfer. - Transfer 100.0 microliter of each
cultures (0 hr timepoint)
sample to 96 well microplateplate 1
in the wells indicated in the plate layout. Maintain at 4.0 degree Celsius during transfer. - Transfer 100.0 microliter of
LB Broth+chloramphenicol
sample to wells A1:H1, A10:H10, A12:H12 of 96 well microplateplate 1
. Maintain at 4.0 degree Celsius during transfer. These samples are blanks.
- Measure 0 hr absorbance timepoint of
plate 1
at 600.0 nanometer. - Measure 0 hr fluorescence timepoint of
plate 1
with excitation wavelength of 488.0 nanometer and emission filter of 530.0 nanometer and 30.0 nanometer bandpass - Cover
plate 1
samples in 96 well microplate with your choice of material to prevent evaporation. - Incubate all
back-diluted culture
samples for 6.0 hour at 37.0 degree Celsius at 220.0. - Incubate all
plate 1
samples for 6.0 hour at 37.0 degree Celsius at 220.0. - Hold all
cultures (0 hr timepoint)
samples at 4.0 degree Celsius. This will inhibit cell growth during the subsequent pipetting steps. - Hold all
plate 1
samples at 4.0 degree Celsius. This will inhibit cell growth during the subsequent pipetting steps. - Provision 16 x 1.5 mL microfuge tubes to contain
6hr timepoint
- Provision a 96 well microplate to contain
plate 2
- Hold
6hr timepoint
at 4.0 degree Celsius. This will prevent cell growth while transferring samples. - Hold
plate 2
at 4.0 degree Celsius. - Transfer 1.0 milliliter of each of 16
back-diluted culture
samples to 1.5 mL microfuge tube containers to contain a total of 166hr timepoint
samples. Maintain at 4.0 degree Celsius during transfer. - Transfer 100.0 microliter of each
6hr timepoint
sample to 96 well microplateplate 2
in the wells indicated in the plate layout. Maintain at 4.0 degree Celsius during transfer. - Transfer 100.0 microliter of
LB Broth+chloramphenicol
sample to wells A1:H1, A10:H10, A12:H12 of 96 well microplateplate 2
. Maintain at 4.0 degree Celsius during transfer. These are the blanks. - Perform a brief centrifugation on 96 well microplate containing
plate 1
samples. This will prevent cross-contamination when removing the seal. - Remove the seal from 96 well microplate containing
plate 1
samples. - Measure 6 hr absorbance timepoint of
plate 1
at 600.0 nanometer. - Measure 6 hr fluorescence timepoint of
plate 1
with excitation wavelength of 485.0 nanometer and emission filter of 530.0 nanometer and 30.0 nanometer bandpass - Measure 6 hr absorbance timepoint of
plate 2
at 600.0 nanometer. - Measure 6 hr fluorescence timepoint of
plate 2
with excitation wavelength of 485.0 nanometer and emission filter of 530.0 nanometer and 30.0 nanometer bandpass - Import data for
baseline absorbance of culture (day 2) measurements of culture (day 2)
,0 hr absorbance timepoint measurements of plate 1
,0 hr fluorescence timepoint measurements of plate 1
,6 hr absorbance timepoint measurements of plate 1
,6 hr fluorescence timepoint measurements of plate 1
,6 hr absorbance timepoint measurements of plate 2
,6 hr fluorescence timepoint measurements of plate 2
into provided Excel file.