RS-232 to TCP/IP bridge for old SCPI instruments

Make old instruments with RS-232 SCPI ports accessible over the network.

Some old instruments, like the Agilent E36xxA family of power supplies, have only RS-232 and GPIB ports for remote control. They lack LAN ports for TCP/IP remote control. Use a RaspberryPi or similar SBC as an RS-232 to TCP/IP bridge to make these devices accessible over the network. This method tunnels all the traffic over a TCP socket. pyvisa can then connect to them using the :SOCKET protocol, such as:

TCPIP0::192.168.2.238::6000::SOCKET                                             

Hardware

Requirements:

  • Raspberry Pi or similar with free USB port(s)
  • Old SCPI instrument(s) with RS-232 serial comms interface
  • FTDI USB-to-serial adapter(s)
  • RS-232 cables

Installation

Install prerequisites:

sudo apt-get install at ncat

Plug in the FTDI USB serial ports. Do this one at a time and identify the serial numbers, such as FTXQI63U. Make note of these serial numbers and which instruments they are connected to. Connect RS-232 cables to the instruments. You might possibly need NULL modem cables.

$ sudo dmesg |grep -E 'FTDI|SerialNumber'
[    5.338991] usb 1-1.2: Manufacturer: FTDI
[    5.346826] usb 1-1.2: SerialNumber: FTXQI63U
[   25.973981] usbserial: USB Serial support registered for FTDI USB Serial Device
[   25.974567] ftdi_sio 1-1.2:1.0: FTDI USB Serial Device converter detected
[   26.170823] usb 1-1.2: FTDI USB Serial Device converter now attached to ttyUSB0

Install these files on RaspberryPi.

$ sudo cp start_visa_redirect.sh /bin  
$ sudo cp 52-ftdi.rules /etc/udev/rules.d/   
$ sudo chmod +x /bin/start_visa_redirect.sh                                                                                                                                                                                                                                                 

Edit /etc/udev/rules.d/52-ftdi.rules:

  • Change the ATTR{serial} numbers to match the serial numbers of your USB devices.
  • Change the SYMLINK and RUN names to something relevant for your equipment.
  • If you want, change the TCP/IP ports from the default 6000 and 6001 to something else.

If your instrument uses something other than 9600-8-n-1, edit start_visa_redirect.sh and change the baud rate in the stty commmand.

Refresh the udev daemon:

$ sudo udevadm control --reload-rules && udevadm trigger

The scripts should now start running. They will log a message to syslog when starting:

$ tail /var/log/messages
Jul 13 05:10:24 raspberrypi root: Starting TCP/IP socket on port 6001 for serial port /dev/power_supply_e3631a

You should see new processes running:

$ ps auxw
root       425  0.0  0.3   3272  1576 ?        SN   05:10   0:00 /bin/bash /bin/start_visa_redirect.sh /dev/power_supply_e3632a 6000
root       429  0.0  0.3   3272  1640 ?        SN   05:10   0:00 /bin/bash /bin/start_visa_redirect.sh /dev/power_supply_e3631a 6001
root       451  0.0  0.6   9392  3020 ?        SN   05:10   0:00 ncat -k -l 6000
root       452  0.0  0.6   9392  3024 ?        SN   05:10   0:00 ncat -k -l 6001

Configure the instrument serial port (default is 9600-8-n-1). Unforunately the DTR/DSR flow control method is unsupported, so we have to be careful when sending commands to avoid overflowing the buffers.

Usage

On the remote computer, use pyvisa to access the instrument. See example.py for a simple script to query the instrument. You'll need to edit the IP address in the file to match your Raspberry Pi.

$  python example.py
HEWLETT-PACKARD,E3632A,0,1.2-5.0-1.0

You will have to add a short delay after each command (0.5 second or maybe longer) to avoid overflowing the instrument's buffers. For example:

def write(msg):                                                       
    # Communication fails unless we delay after sending each command..      
    ret = instrument.write(msg)                                                
    time.sleep(0.5)                                                         
    return ret    

This is not ideal, and it might not work if the network link has variable latency. At least on a LAN, it has worked well enough for me. Anyone have a better idea? Maybe rewire the port to connect DTR/DSR flow control lines to RTS/CTS and enable crtscts in stty?